Abstract
A new method for detecting low levels of growth promoters in bovine meat was developed with the following goal: easy, fast and sensitive analysis of a wide range of compounds, with reduced consumption of chemicals and disposables. Several classes of growth promoters were included, i.e. resorcylic acid lactones (RALs) and steroids, the latter including corticosteroids and gestagens. For sample treatment, 0.5 g of homogenised bovine meat was simultaneously disrupted and extracted in a bead-ruptor machine. The organic extraction solvent was further processed by solid-phase extraction (SPE) clean-up using 96-Well Oasis® HLB Plates. Six SPE washing steps were applied to remove matrix compounds after which the growth promoters were eluted and analysed using UHPLC-MS/MS. To achieve lower detection levels and to reduce LC-solvent consumption, a narrow-bore column with an internal diameter of 1 mm was used, instead of the conventional 2.1 mm. During analysis, the mass spectrometer was operated in negative and positive ionisation mode (ion switching). The newly developed method was validated according to the Commission Decision 2002/657. The results demonstrate that the method meets the criteria as established in this Commission Decision. The precision of the method for exogenous steroids varies between 85 and 115%, the CCα for the compounds ranges from 0.1–0.9 μg kg−1 and the expanded measurement uncertainty was lower than 36%. Compared to our current in-house methods with analysis times of 2 days for a maximum of 24 samples, the new method offers improved sample throughput (96 samples in less than 24 h) and lower detection limits.
Similar content being viewed by others
Introduction
In Europe, the use of growth promoters in animal production is strictly forbidden (Council Directive 96, 1996; Directive C, 1996). A zero-tolerance policy is applied for these substances, which means in practice that the analysis has to be performed at the lowest concentration possible (Committee ES, 2012). There are several classes of growth promoters including stilbenes, steroids and beta-agonists. There is ongoing debate on the risks for humans when steroids are used as growth promoters (Committee ES, 2012). Endocrine disruption is regarded as one of the main health issues related to the use of hormonal growth promoters. Therefore, it is expected that the EU ban on using growth promoters for fattening will not be lifted in the near future which makes it necessary to improve current control methods so that a larger number of compounds can be detected at low concentrations in food (Koster et al., 2011). Routine analysis of meat samples is mainly carried out using methods based on LC-MS/MS or GC-MS/MS. However, in order to comply with the latest views on acceptable concentrations described in the EFSA opinion (Panel, 2013), a decrease of the limits of detection of current analytical methods is needed.
Current routine methods of analysis of growth promoters in meat typically consist of processing 2–5-g portions of meat by either a digestion step with the enzyme subtilisin (Blasco et al., 2007; Daeseleire et al., 1991), an ultrasonic disruptor (Kinsella et al., 2009) and lyophilisation/freeze drying, followed by supercritical fluid extraction (Marchand et al., 2000; Stolker et al., 1998), defatting (Barkatina et al., 2001; Stolker & Brinkman, 2005), several LLE (liquid-liquid extraction) steps or a few SPE (solid-phase extraction) steps (Bovee et al., 2009; Deceuninck et al., 2013), and sometimes even HPLC fractionation (van Ginkel et al., 1992), followed by mass spectrometric detection. Some methods include a deconjugation step (Malone et al., 2009), although there is a negligible amount of phase II conjugates present in meat (Maume et al., 2001). Often, significant background signals due to interfering compounds such as proteins, salts and fatty acids (Antignac et al., 2005) are observed. Typically, processing 24 samples takes approximately two working days with our current methods, using large amounts of chemicals and organic solvents. Recent literature demonstrates that sample clean-up can in principle be more efficient. For example, smaller aliquot of samples of meat can be processed to reduce matrix effects, so less matrix compounds will be introduced during the final analysis (Stahnke et al., 2012). Also, the use of 96-wells format increase the efficiency of sample clean-up (O’Brien et al., 2009; Pinel et al., 2008). In a recent paper of Luo et al. (Luo et al., 2016), the use of hollow-fibre liquid-phase microextraction (HF-LPME) was combined in a 96-wells format to extract 16 steroids from urine, milk and plasma in only a few hours. Another fast method is based on a modified Quick Easy Cheap Effective Rugged Safe (QuEChERS) approach (Attalah et al., 2016) to extract four steroids from meat. This is a fast method and applicable in many laboratories, although sensitivity and number of compounds need to be improved to widen the applicability of the method. Besides optimising sample clean-up procedures, efficiency can also be increased by the application of full-scan accurate mass measurement (HRAM) instead of triple-quad measurements as described by Lynn Vanhaecke et al. (Vanhaecke et al., 2013). In this study, it is demonstrated that the performance of HRAM rivals the performance of triple-quad measurements in combination with a traditional clean-up. The use of HRAM opens the door to the detection of an unlimited number of compounds and it can pursue the perspective of untargeted strategies and retrospective analyses.
In spite of the advances made during the past years, the current described methods for the analysis of steroids in samples of muscle tissue lack the sensitivity which is necessary to comply with the latest EFSA opinion (Panel, 2013). Also, they are relatively time consuming, detect single classes of compounds and are environmental unfriendly. The objective of this study was to redesign our current methods in such a way that the sensitivity is increased, the workload is reduced and the use of chemicals is decreased by miniaturisation. Although not the explicit goal of this study, the principles as described by Gałuszka et al. (Gałuszka et al., 2013) to make analytical methods ‘greener’ were adopted by implementing miniaturisation, resulting in reduction of the amount of chemicals. The developed method was validated by measuring the precision, linearity, decision limit (CCα), detection capability (CCβ) and measurement uncertainty according to the guidelines established by the European Union (Commission E, 2002). Method optimisation and validation are both described.
Experimental
Solvents and Reagents
Methanol (MeOH), water, acetonitrile (ACN) and formic acid (FA) (all ULC/MS grade) were obtained from Actu-All Chemicals (Oss, The Netherlands); ammonia (25%), acetic acid (100%) and ethanol from Merck Millipore (Darmstadt, Germany); and ammonium formate from Sigma-Aldrich (St. Louis, MO, USA). Milli-Q water was prepared using a Milli-Q system at a resistivity of at least 18.2 MΩ cm−1 (Merck Millipore).
Standards
ADD (androsta-1,4-diene-3,17-dione), betamethasone, 17ß-boldenone, clobetasol, isoflupredone, medroxyprogesterone-acetate, methylboldenone, 17α-methyltestosterone, 17α-nortestosterone, 17β-nortestosterone, 17α-testosterone, 17β-testosterone and 17ß-trenbolone were obtained from Steraloids (Newport, USA); chlormadinone-acetate, dexamethasone, flumethasone, megestrolacetate, melengestrolacetate, prednisolone, methylprednisolone, triamcinolone-acetonide, α-zearalenol, ß-zearalenol, zearalenone, α-zeranol and ß-zeranol from Sigma-Aldrich (St. Louis, USA); α-trenbolone and 16β-OH-stanozolol from NMI (Australia, Sydney); 17α-boldenone from the EURL (Wageningen, the Netherlands); stanozolol from Sequoia Research Products (Pangbourne, UK) and zearalanone from TRC (Toronto, Canada).
The deuterium-labelled standards 17α-boldenone-d3, 17ß-boldenone-d3, megestrolacetate-d3, methylboldenone-d3, medroxyprogesterone-d3, 17α-methyltestosterone-d3, 17β-nortestosterone-d3, 16α-OH-stanozolol-d3, 17α-trenbolone-d3, 17ß-trenbolone-d3, triamcinolone-acetonide-d6, α-zearalanol-d4/ß-zearalanol-d4, α-zearalenol-d4 and ß-zearalenol-d4 were obtained from the EURL (Wageningen, the Netherlands); melengestrolacetate-d3, prednisolone-d8 (d7 major), zearalanone-d6 and zearalenone-d6 from TRC (Toronto, Canada); dexamethasone-d4 from Cachesyn (Mississauga, Canada); stanozolol-d3 from Cerilliant (Round Rock, USA) and 17β-testosterone-d3 from NMI (Australia, Sydney).
Materials
The extraction plate manifold, Oasis HLB 96-well plate (60-mg sorbent per well, 60-μm particle size), 96-well collection plates (2 ml), 96-well PTFE/silicone seal with pre-slit, as well as the Acquity UPLC BEH C18 column (1.7 μm, 1.0 × 100 mm) were purchased from Waters (Milford, MA, USA). The bead-ruptor machine and polypropylene reaction tubes were purchased from Omni International (NW Kennesaw, USA), and beads from BioSpec Products (Bartlesville, USA), the evaporator for 96-well plates from Biotage (Uppsala, Sweden) and the centrifuge from Heraeus (Hanau, Germany). The UHPLC-MS/MS system consisted of a Waters ACQUITY UPLC® I-Class System connected to a Waters triple quadrupole mass spectrometer Xevo TQS with an ESI interface (Milford, MA, USA). For this study, only bovine meat was used.
Sample Extraction and Clean-up
Comparisons of Digestions and Primary Extractions from Bovine Meat
To investigate whether the bead-ruptor extracts the components efficiently from the meat, several common extraction procedures (Blasco et al., 2007; Daeseleire et al., 1991; Kinsella et al., 2009; Barkatina et al., 2001) were tested and the outcomes were compared. With 1 g of meat, the following extraction procedures were tested: 2 ml TRIS buffer, pH 9.5, with 10 mg subtilisin for 16 h; 2 ml acetate buffer, pH 5.2, with 10 μl Helix SMP for 16 h; 2 ml phosphate buffer, pH 7.8, with 10 μl beta-glucuronidase for 16 h; 2 ml TRIS buffer, pH 9.5; 2 ml acetate buffer, pH 5.2, for 16 h; 2 ml phosphate buffer, pH 7.8, for 16 h; ultrasonic disruptor in 2 ml water for 30 s; bead-disruption in 2 ml 60/40 acetonitrile/H2O for 2 min. The eight different digested meat samples were then subjected to visual inspection by microscopic examination (× 200), examining the amount of intact muscle tissue, the nuclei and the general status of the meat. Three types of sections were made using 100 μl of the final digest: without addition of modifiers (native) and with addition of haematoxylin or Newmans containing TBS, as colouring methods. As visual inspection of the treated meat samples might be difficult and subjective, the extraction efficiency also was determined by measuring the concentration of endogenous 17α/β-testosterone for each tested primary extraction technique. After extraction, all extracts were processed with the method as described in the next section, using a high throughput 96-wells purification procedure.
Primary Extraction from Meat by Bead-Disruption
To 500 mg of homogenised bovine meat sample or standard, 2 μg kg−1 of (isotope-labelled) internal standard mixture (see Table 1 for compounds included in the internal standard mixture) was added. A calibration curve was prepared by spiking blank meat samples with standards (see Table 1 for compounds included in the standard mixture) at concentrations of 0, 0.2, 0.5, 1.0, 2.5, and 5.0 μg kg−1 and 1 ml of extraction solvent, i.e. 60% acetonitrile, were added. Bead-disruption was carried out on the bead-ruptor machine using the following program: speed, 5.65 m s−1; cycle time, 0.75 min; number of cycles, 2 and pause dwell, 0.50 min. Subsequently, the tubes were centrifuged for 10 min at 3452 g and the supernatant was transferred into a clean glass tube, while avoiding transferring the fat layer as much as possible. To dilute the transferred supernatant to a final content of 20% acetonitrile (to prevent the loss of analytes during application onto the Oasis HLB sorbent), 2 ml of water was added.
SPE 96-wells Purification
An Oasis HLB 96-well plate was first conditioned with 1 ml methanol, followed by 1 ml water. Samples were transferred in two portions to the SPE 96-well plate and each well was subsequently washed with 1 ml water. Next, six consecutive washing steps of 1 ml each were applied using different solvents, i.e. 60% methanol in water/2% formic acid, 20% acetonitrile in water/2% formic acid, 20% acetonitrile in water, 10% acetonitrile in water/2% ammonia, 50% methanol in water/2% ammonia, and 60% methanol in water. Elution of analytes was performed with 1 ml acetonitrile, and the eluate was collected in a 96-well collection plate already containing 20 μl DMSO per well as a keeper. The plate was placed in a 96-well plate evaporator and dried until only the DMSO was left. Then, 30 μl of 10% methanol was added and the plate was gently shaken. For analysis, 10 μl of the final residue was injected into the UHPLC-MS/MS system.
UHPLC-MS/MS Analysis
Chromatographic separation was performed on an UPLC BEH C18, 100 mm × 1.0 mm, 1.7-μm analytical column. Mobile phase A consisted of water/acetonitrile/ammonium formate 1 M/formic acid 900/100/2/0.02 (v/v/v/v), and mobile phase B of water/acetonitrile/ammonium formate 1 M/formic acid 100/900/2/0.02 (v/v/v/v). Column temperature was kept at 60 °C, vial tray temperature was 20 °C, flow rate was 0.15 ml/min and injection volume was 10 μl. All analytes were eluted using the following linear gradient: 0–0.2 min 80% A, 0.2–3.2 min 70% A, 3.2–7.5 min 50% A, 7.5–8.6 min 0% A and 8.6–10.0 min 80% A.
The MS system switched between positive and negative ion ionisation during analysis, and the capillary voltage was set on 3.0 kV, with a cone gas flow of 150 l per hour. The source temperature was 150 °C, and the desolvation temperature was 400 °C. The flow of the desolvation gas was 800 l per hour. The LM 1 resolution was 3.0 in ESI-positive and 2.8 in ESI-negative mode. The HM 1 and HM 2 resolution was 15. The ion energy 1 was set to 1.0 in ESI-positive and 1.9 in ESI-negative mode. The ion energy 2 was set to 1.0 in ESI-positive and 1.3 in ESI-negative mode. Table 1 gives an overview of the measured transitions, and the associated cone voltages for each compound are included.
Data Processing
UHPLC-MS/MS data was automatically processed using MassLynx V4.1 software from Waters. All peak integrations were manually checked and baseline corrected if necessary. Recovery losses during sample clean-up were corrected by the use of suitable internal standard compounds as denoted in Table 1.
Method Validation
The described method was validated as a quantitative screening method according to the Commission Decision 2002/657/EC (European Commission 2002) (Commission E, 2002) and was performed for all compounds. For all compounds, the CCα, CCß, precision, repeatability and measurement uncertainty (MU) were determined. The decision limit (CCα) and detection capability (CCß) were determined by spiking at three different concentration levels (1, 2 and 3 μg kg−1) in different blank meat samples. Eight different bovine meat samples were obtained from routine control programs in the Netherlands. One of these meat samples was used for the matrix-matched standard calibration curve, the other meat samples for the validation. Each individual concentration was fully processed in seven-fold as if it was an unknown sample and this set of samples was analysed on 1 day. From the obtained calibration curve, the y-intercept and slope were calculated (y = ax+b) and the CCα, CCß and MU were calculated according to ISO11843 (Van Loco & Beernaert, 2003). For maximum residue level (MRL) compounds, the CCα was calculated as MRL plus 1.64 times the standard deviation of the within-laboratory reproducibility. CCβ MRL was calculated as the value of the decision limit plus 1.64 times the standard deviation of the within-laboratory reproducibility. The expanded MU was calculated using a coverage factor of k = 2 and a confidence level of 95%. The repeatability was determined from the obtained dataset using the approach as described in ISO 5725 (Guide I. 5725–1: accuracy (trueness and precision) of measurement methods and results. Part 1: general principles and definitions, 1994). The overall precision was calculated by taking the square root of the sum of the squares of the precision on each spiking level divided by 3. The stability of the extracts was assessed by reinjecting samples after 1-week storage at − 20 °C.
Results and Discussion
Optimisation of the Digestion and Primary Extraction from Meat
The primary digestion can be performed in three different ways: chemical, enzymatic, or mechanical. These three digestive procedures were tested, and the remaining muscle fibres were microscopically examined. The chemical extraction with 2 ml phosphate buffer, pH 7.8, did almost not affect the meat, i.e. the extracts still contained intact muscle fibres. Chemical extraction at mild acid (2 ml acetate buffer, pH 5.2) or alkali conditions (2 ml TRIS buffer, pH 9.5) showed more destruction of the muscle tissue, but the residue still contained visible cell nuclei. Enzymatic digestion with 2 ml phosphate buffer, pH 7.8, with 10 μl beta-glucuronidase was not very effective either; the residue contained intact muscle fibres. Enzymatic digestion with 2 ml TRIS buffer, pH 9.5, with 10 mg subtilisin and 2 ml acetate buffer pH 5.2 with 10 μl Helix SMP was effective, and most of the meat was digested. Mechanical digestion was carried out by means of a bead-ruptor machine. This bead-disruptor is specifically designed for grinding, lysing and homogenising biological samples prior to extraction. In ‘bead beating’, a large number of minute glass, ceramic or steel beads are vigorously agitated by shaking. Disruption occurs by the crushing action of the beads as they collide with the tissue. Another mechanical digestion technique is by use of an ultrasonic disruptor whereby a high-frequency energy disrupts the cells. Microscopic examination showed that the bead-disruption in 2 ml 60/40 acetonitrile/H2O and by use of an ultrasonic disruptor in 2 ml water both contained some visible cell nuclei and remnants of muscle fibres the rest of tissue was digested. Figure 1 shows microscopic photographs of intact meat and meat treated by bead-disruption in acetonitrile/water.
As an alternative approach for determination of the extraction efficiency, the levels of endogenous concentrations of 17α- and 17β-testosterone were determined. Figure 2 shows the concentrations for each extraction technique tested (a–h). From Fig. 2, it can be concluded that the highest concentrations were obtained by use of the bead-ruptor (h) for both testosterone isomers compared to the other digestion and extraction techniques (a–f). Based on the microscopic results, it was expected that the enzymatic digestion with 2 ml TRIS buffer, pH 9.5, with 10 mg subtilisin and 2 ml acetate buffer pH 5.2 with 10 μl Helix SMP would have given the highest yield since the meat was completely digested. The lower yield of the two enzymatic digestions compared to that of the mechanical digestion could be explained by the fact that enzymatic digestion takes 16 h at elevated temperatures resulting in losses of the compounds. The most efficient digestion, timewise, was using the bead-ruptor since 24 samples can be simultaneously processed in 2 min resulting in minimal losses during digestion as depicted in Fig. 2.
Optimization of a 96-well SPE Extraction Method
For a more efficient workflow and to increase the sample throughput, solid-phase extraction using single SPE columns was converted to a 96-well plate format. To obtain the highest possible recoveries, the influence of the 96-well clean-up materials on the analyte recoveries was investigated. It turned out that large losses of compounds occurred at the solvent-evaporation step in the 96-well collection plates. Most probably, compounds were adsorbed to the plastic material of the 96-well plate, hindering reconstitution. These recovery losses were plate and brand independent. To prevent these losses, a small volume of dimethylsulfoxide (DMSO) was added before the evaporation step, i.e. DMSO serving as a ‘keeper’ solvent. After addition of DMSO as the keeper, losses were minimal (see supplementary information).
Due to the efficient workflow in the 96-well format, it is easy to test numerous and different washing steps. The following washing solvents were tested whereby the organic solvent percentage was increased from 10 to 90% in steps of 10%: methanol/water/formic acid (2%); methanol/water/ammonia (2%); methanol/water; acetonitrile/water/formic acid (2%); acetonitrile/water/ammonia (2%) and acetonitrile/water. For each compound, the corresponding elution profile was constructed. Figure 3 shows an example of such an elution profile.
Next, for each solvent mixture, the critical breakthrough washing condition was selected. That is, the composition with the lowest percentage of organic modifier determines the critical breakthrough condition for that washing solvent (washing step) in the whole method. The following critical washing conditions were determined: 60% in methanol/water/formic acid; 50% in methanol/water/ammonia; 60% in methanol/water; 20% in acetonitrile/water/formic acid; 20% in acetonitrile/water/ammonia and 20% in acetonitrile/water. After determination of these critical compositions, an experimental design was constructed to test the optimal number and order of the selected washing solvents (Table 2).
The chromatograms for all combinations were visually evaluated, for some of the combinations interfering peaks were visible from the meat extracts, making integration more difficult which will eventually result in unreliable quantification. It was noted that the cleanest chromatograms were obtained using a combination of a protic and an aprotic organic solvent, e.g. methanol and acetonitrile. So, combinations E and H (only protic) and F and I (only aprotic) were less effective in removing interfering peaks. Washing combinations with only neutral (combination B), only acid (combination C) or only alkaline (combination D) washing solvents were also unsuccessful in removing all interfering peaks. The optimum number and order of solvents were found to be a combination of acid and alkaline protic and aprotic solvents. At the end, the optimal washing condition consisted of six different solvents in the following order: 60% methanol/formic acid, 50% methanol/ammonia, 60% methanol, 20% acetonitrile/formic acid, 20% acetonitrile/ammonia and 20% acetonitrile. Figure 4 shows that this comprehensive washing procedure (Table 2 combination G) produced lower background signals compared to the two washing steps method with two different solvents (Table 2 combination J) and also resulted in minimal losses during sample clean-up for most of the compounds.
Optimization of the UPLC Conditions
To increase the sensitivity and reduce the organic solvent usage, a column with an internal diameter of 1.0 mm was used instead of the commonly used 2.1 mm. From a theoretical point of view, the sensitivity should increase quadratic with linear reduction of the column diameter when the same mass amount of sample is injected. In this case, switching from a 2.1-mm column to a column with an ID of 1.0 mm would theoretically result in a factor of 4 increase in sensitivity. It should be noted that this factor of 4 can only be reached if all LC connections are made properly and the dead volume is reduced to a minimum. Figure 5 gives an example of α/ß-nortestosterone whereby the same amount was injected on a 2.1- and 1.0-mm column. Comparing the signal to noise ratio, but also the peak height, demonstrates that there is indeed an increase in sensitivity, in general, the sensitivity of this column was approximately a factor of 4 better. In addition, the amount of organic solvents used was reduced, as the flow was decreased from 0.6 ml/min on a 2.1-mm column to 0.15 ml/min on a 1-mm column, this is a 75% reduction of organic solvent.
Validation
The method was validated as a screening method according to the guidelines set in the Commission Decision 2002/657 (Commission E, 2002). The validation results for precision, CCα, CCß, linearity, recovery and measurement uncertainty are presented in Table 3. The method was validated for all compounds as if they were forbidden substances. However, this is not the case for dexamethasone, betamethasone, prednisolone and methylprednisolone which are permitted for therapeutic purposes. There are MRLs established for these compounds. The MRL for dexamethasone and betamethasone is 0.75 μg kg−1; for prednisolone, 4 μg kg−1 and for methylprednisolone, 10 μg kg−1. The CCα and CCß values were calculated at the MRL as well (see Table 3). As meat contains certain endogenous hormones or unintended contaminants, it is impossible to determine the performance characteristics for these compounds. During validation, 17α-testosterone (n = 7, average concentration 0.23 μg kg−1) and 17ß-testosterone (n = 7, average concentration 1.45 μg kg−1) were detected in the blank samples. From literature, it is also known that prednisolone (Arioli et al., 2010; De Clercq et al., 2015), α/ß-zearalenol and zearalenone originating from Fusarium can be found in blank samples (Kennedy et al., 1998; Seeling et al., 2005; Price & Fenwick, 1985) as well as 17α/ß-nortestosterone (de Brabander et al., 1994). In the blank samples used in this validation study, these endogenous compounds were not detected, except for 17α/ß-testosterone. This results in higher measurement uncertainties, in particular for 17α/ß-testosterone. Therefore, it cannot be neglected that the validation results could be influenced by background concentrations of these compounds. 17α/ß-Testosterone, prednisolone, methylprednisolone and α/ß-zearalenol have higher CCα values than 1 μg kg−1 (Table 3). All other growth promoters have a CCα in the range of 0.1–0.9 μg kg−1 (Table 3). The measurement uncertainty for these growth promoters is lower than 36%. The precision for these exogenous compounds is within the range of 85–115%, which is considered as acceptable for the intended use of this method and in compliance with the requirements as described in the EU Commission Decision 2002/657/EC. All matrix-matched calibration curves resulted in a R2 > 0.99 in the range from 0.2 to 5 μg kg−1 except for ß-zearalanol, α-zearalenol, prednisolone and methylprednisolone. The overall sensitivity of these compounds is lower compared to that of the other compounds resulting in a lower correlation coefficient. The recovery for most compounds is between 30 and 55%, which is acceptable at residue levels. The recoveries for the resorcylic acid lactones, prednisolone and isoflupredone are low, 8–20%. The poor recovery of these compounds is due to the extensive washing procedure. The specificity was tested by the analysis of eight different blank meat samples: apart from the naturally occurring compounds, no chromatographic peaks were present at the expected retention time of the individual checked growth promoters. This indicates the absence of interfering compounds that might result in false positive outcomes. The selectivity of the MRM transitions was assessed according to the description of Berendsen et al. (Berendsen et al., 2013). All measured transitions were found to be selective. Sample extracts were found to be stable for at least 1 week.
The results of the validation study confirm that the newly developed 96-well high throughput clean-up procedure in combination with the optimised UHPLC-MS/MS analysis results in a greener method for the fast and broad detection of growth promoters in meat, showing acceptable performance characteristics that fulfil the performance criteria as established in CD 2002/657/EC.
Conclusions
An easy, fast and sensitive method for broad analysis of different growth promoters, with reduced consumption of chemicals and disposables, was developed. The compounds analysed with this method are different classes of steroids, corticosteroids, gestagens and resorcylic acid lactones. So far, no other published method is able to analyse these different classes of compounds in meat in a single method at levels below 1.0 μg kg−1 as recommended by EFSA (Panel, 2013). Overall, this new method offers an increased sample throughput: sample pre-treatment, analysis of 96 samples and analysis of calibration and control samples in less than 24 h. As a result of the reduction in analysis time and consumables needed for sample clean-up, the costs for analysis are substantially reduced. The precision of the method for exogenous steroids varies between 85 and 115%, the CCα for the compounds ranges from 0.1–0.9 μg kg−1 and the expanded measurement uncertainty was lower than 36%. The validation proved that the method is reliable and sensitive. The method can be used as a routine screening method to detect abuse of growth promoters in cattle fattening.
Furthermore, the method can be extended to include other species and matrices. An interesting future application could be the use of this clean-up method in combination with full-scan accurate mass spectrometric measurement to obtain a true generic method which is capable of analysing an unlimited number of growth promoters in meat.
References
Antignac J-P, de Wasch K, Monteau F, De Brabander H, Andre F, Le Bizec B (2005) The ion suppression phenomenon in liquid chromatography–mass spectrometry and its consequences in the field of residue analysis. Anal Chim Acta 529(1):129–136. https://doi.org/10.1016/j.aca.2004.08.055
Arioli F, Fidani M, Casati A, Fracchiolla ML, Pompa G (2010) Investigation on possible transformations of cortisol, cortisone and cortisol glucuronide in bovine faecal matter using liquid chromatography-mass spectrometry. Steroids 75(4–5):350–354. https://doi.org/10.1016/j.steroids.2010.01.016
Attalah E, Nasr YS, El-Gammal HA, Nour El-Dien FA (2016) Optimisation and validation of a new analytical method for the determination of four natural and synthetic hormones using LC-ESI-MS/MS. Food Addit Contam Part A Chem Anal Control Expo Risk Assess 33(10):1545–1556. https://doi.org/10.1080/19440049.2016.1227878
Barkatina E, Volkovich S, Venger O, Murokh V, Kolomiets N, Shulyakovskaya O (2001) Simultaneous determination of diethylstilbestrol, testosterone, and 17β-estradiol residues in meat and meat products using gas–liquid chromatography. J Anal Chem 56(8):740–743
Berendsen BJ, Stolker LA, Nielen MW (2013) The (un) certainty of selectivity in liquid chromatography tandem mass spectrometry. J Am Soc Mass Spectrom 24(1):154–163. https://doi.org/10.1007/s13361-012-0501-0
Blasco C, Van Poucke C, Van Peteghem C (2007) Analysis of meat samples for anabolic steroids residues by liquid chromatography/tandem mass spectrometry. J Chromatogr A 1154(1–2):230–239. https://doi.org/10.1016/j.chroma.2007.03.090
Bovee TF, Bor G, Heskamp HH, Lasaroms JJ, Sanders MB, Nielen MW (2009) Validation and application of a yeast bioassay for screening androgenic activity in calf urine and feed. Anal Chim Acta 637(1–2):225–234. https://doi.org/10.1016/j.aca.2008.06.047
de Brabander HF, van Hende J, Batjoens P, Hendriks L, Raus J, Smets F, Pottie G, van Ginkel L, Stephany RW (1994) Endogenic nortestosterone in cattle? Analyst 119(12):2581–2585. https://doi.org/10.1039/an9941902581
Commission E. Commission Decision EC 2002/657 of 12 August 2002 implementing Council Directive 96/23/EC concerning the performance of analytical methods and the interpretation of results. Off J Eur Communities L 2002;221
Committee ES. Scientific opinion on exploring options for providing advice about possible human health risks based on the concept of threshold of toxicological concern (TTC). EFSA J 2012;10(7)
Council Directive 96 (1996) 22/EC of 29 April 1996 concerning the prohibition on the use in stockfarming of certain substances having a hormonal or thyrostatic action and of ß-agonists, and repealing directives 81/602/EEC, 88/146/EEC and 88/299/EEC. Official Journal L 125(23/05):0003–0009
Daeseleire E, De Guesquiere A, Van Peteghem C (1991) A simple procedure for the enzymic digestion of edible tissues prior to detection of drug residues. Z Lebensm Unters Forsch 192(2):105–107. https://doi.org/10.1007/BF01202620
De Clercq N, Van Meulebroek L, Vanden Bussche J, Croubels S, Delahaut P, Vanhaecke L (2015) The impact of stress on the prevalence of prednisolone in bovine urine: a metabolic fingerprinting approach. J Steroid Biochem Mol Biol 154:206–216. https://doi.org/10.1016/j.jsbmb.2015.08.026
Deceuninck Y, Bichon E, Monteau F, Dervilly-Pinel G, Antignac J, Le Bizec B (2013) Fast and multiresidue determination of twenty glucocorticoids in bovine milk using ultra high performance liquid chromatography–tandem mass spectrometry. J Chromatogr A 1294:76–86. https://doi.org/10.1016/j.chroma.2013.04.019
Directive C (1996) Council directive 96/23/EC of 29 April 1996 on measures to monitor certain substances and residues thereof in live animals and animal products and repealing directives 85/358/EEC and 86/469/EEC and decisions 89/187/EEC and 91/664/EEC. OJ EC L 125:10–31
Gałuszka A, Migaszewski Z, Namieśnik J (2013) The 12 principles of green analytical chemistry and the SIGNIFICANCE mnemonic of green analytical practices. TrAC Trends Anal Chem 50:78–84
van Ginkel LA, Jansen EH, Stephany RW, Zoontjes PW, Schwillens PL, van Rossum HJ et al (1992) Liquid chromatographic purification and detection of anabolic compounds. J Chromatogr 624(1–2):389–401. https://doi.org/10.1016/0021-9673(92)85690-U
Guide I. 5725–1: accuracy (trueness and precision) of measurement methods and results. Part 1: general principles and definitions. ISO: Geneva; 1994
Kennedy DG, Hewitt SA, McEvoy JDG, Currie JW, Cannavan A, Blanchflower WJ, Elliot CT (1998) Zeranol is formed from Fusarium spp. toxins in cattle in vivo. Food Addit Contam 15(4):393–400. https://doi.org/10.1080/02652039809374658
Kinsella B, O’Mahony J, Malone E, Moloney M, Cantwell H, Furey A et al (2009) Current trends in sample preparation for growth promoter and veterinary drug residue analysis. J Chromatogr A 1216(46):7977–8015. https://doi.org/10.1016/j.chroma.2009.09.005
Koster S, Boobis AR, Cubberley R, Hollnagel HM, Richling E, Wildemann T, Würtzen G, Galli CL (2011) Application of the TTC concept to unknown substances found in analysis of foods. Food Chem Toxicol 49(8):1643–1660. https://doi.org/10.1016/j.fct.2011.03.049
Luo GZ, Li YX, Bao JJ (2016) Development and application of a high-throughput sample cleanup process based on 96-well plate for simultaneous determination of 16 steroids in biological matrices using liquid chromatography-triple quadrupole mass spectrometry. Anal Bioanal Chem 408(4):1137–1149. https://doi.org/10.1007/s00216-015-9213-1
Malone E, Elliott C, Kennedy D, Regan L (2009) Development of a rapid method for the analysis of synthetic growth promoters in bovine muscle using liquid chromatography tandem mass spectrometry. Anal Chim Acta 637(1):112–120. https://doi.org/10.1016/j.aca.2008.09.002
Marchand P, le Bizec B, Gade C, Monteau F, Andre F (2000) Ultra trace detection of a wide range of anabolic steroids in meat by gas chromatography coupled to mass spectrometry. J Chromatogr A 867(1–2):219–233. https://doi.org/10.1016/S0021-9673(99)01114-0
Maume D, Deceuninck Y, Pouponneau K, Paris A, Bizec B, Andre F (2001) Assessment of estradiol and its metabolites in meat. APMIS 109(S103):S365–S372. https://doi.org/10.1111/j.1600-0463.2001.tb05788.x
O’Brien Z, Post N, Brown M, Madan A, Coon T, Luo R, Kohout TA (2009) Validation and application of a liquid chromatography–tandem mass spectrometric method for the simultaneous determination of testosterone and dihydrotestosterone in rat prostatic tissue using a 96-well format. J Chromatogr B 877(29):3515–3521. https://doi.org/10.1016/j.jchromb.2009.08.053
Panel EB (2013) Scientific opinion on the public health hazards to be covered by inspection of meat (bovine animals). EFSA J 11(6):3266
Pinel G, Rambaud L, Cacciatore G, Bergwerff A, Elliott C, Nielen M, Le Bizec B (2008) Elimination kinetic of 17β-estradiol 3-benzoate and 17β-nandrolone laureate ester metabolites in calves’ urine. J Steroid Biochem Mol Biol 110(1):30–38. https://doi.org/10.1016/j.jsbmb.2007.09.024
Price KR, Fenwick GR (1985) Naturally occurring oestrogens in foods—a review. Food Addit Contam 2(2):73–106. https://doi.org/10.1080/02652038509373531
Seeling K, Danicke S, Ueberschar KH, Lebzien P, Flachowsky G (2005) On the effects of Fusarium toxin-contaminated wheat and the feed intake level on the metabolism and carry over of zearalenone in dairy cows. Food Addit Contam 22(9):847–855
Stahnke H, Kittlaus S, Kempe G, Alder L (2012) Reduction of matrix effects in liquid chromatography-electrospray ionization-mass spectrometry by dilution of the sample extracts: how much dilution is needed? Anal Chem 84(3):1474–1482. https://doi.org/10.1021/ac202661j
Stolker AA, Brinkman UA (2005) Analytical strategies for residue analysis of veterinary drugs and growth-promoting agents in food-producing animals—a review. J Chromatogr A 1067(1–2):15–53. https://doi.org/10.1016/j.chroma.2005.02.037
Stolker AAM, Zoontjes PW, van Ginkel LA (1998) The use of supercritical fluid extraction for the determination of steroids in animal tissues. Analyst 123(12):2671–2676. https://doi.org/10.1039/a805127e
Van Loco J, Beernaert H (2003) An alternative method validation strategy for the European Decision 2002/657/EC. Proceedings of Euro Food Chem XII: Strategies for Safe Food 1:91–94
Vanhaecke L, Van Meulebroek L, De Clercq N, Vanden Bussche J (2013) High resolution orbitrap mass spectrometry in comparison with tandem mass spectrometry for confirmation of anabolic steroids in meat. Anal Chim Acta 767:118–127. https://doi.org/10.1016/j.aca.2013.01.009
Funding
This project was financially supported by the Dutch Ministry of Economic Affairs and the European Commission DG Santé.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of Interest
Frederike van Tricht declares that she has no conflict of interest. Martien Essers declares that he has no conflict of interest. Maria Groot declares that she has no conflict of interest. Saskia Sterk declares that she has no conflict of interest. Marco Blokland declares that he has no conflict of interest. Leen van Ginkel declares that he has no conflict of interest.
Ethical Approval
All applicable international, national and/or institutional guidelines for the care and use of animals were followed.
Informed Consent
Not applicable
Electronic Supplementary Material
ESM 1
(DOCX 84 kb).
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
van Tricht, F., Essers, M., Groot, M. et al. A Fast Quantitative Multi-analyte Method for Growth Promoters in Bovine Meat Using Bead-Disruption, 96-well SPE Clean-up and Narrow-Bore UHPLC-MS/MS Analysis. Food Anal. Methods 11, 2206–2217 (2018). https://doi.org/10.1007/s12161-018-1164-7
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12161-018-1164-7