Introduction

The introduction of penicillin to treat bacterial infections has been one of the most important lifesaving medical discoveries of the twentieth century. Today, penicillins are still universally used both in human and veterinary medicine. Unfortunately, indiscriminate use of penicillins and other ß-lactam antibiotics has been blamed for the increase of bacterial resistance to antibiotics. Furthermore, the consumption of food products containing penicillin drug residues may be harmful to allergic consumers [13]. Therefore, the European Union (EU) has established maximum residue levels (MRLs) for most ß-lactam antibiotics (Fig. 1) in milk and animal tissues (see Table 1) and included them in National Residue Monitoring Programs [4]. Penicillins can be administered to livestock in a number of ways (e.g. in feed, water), but only amoxicillin (AMX) has actually been registered both as a feed and water additive (premix 250–500 mg kg−1). Medication in feed is often preferred due to the low stability of penicillins in aqueous solutions [5]. Analysis and control of antibiotics in feedstuffs have become an important issue as feedstuffs can only be prepared under specific conditions as stated in Council Directive 90/167/EEC [6].

Fig. 1
figure 1

Structures of penicillins, derivatised penicillins and derivatised internal standards

Table 1 MRLs and target values for penicillin residues in animal tissue, milk and feed

Over the years, many methods have been described for the analysis of penicillins in various matrices including animal feeds. Unfortunately, ß-lactams are known for their rapid degradation during sample preparation. The use of organic solvents and acids in extraction and chromatography has been implicated in the degradation of penicillins [79]. Losses may also occur during storage (milk [10], meat [11] and feed [5]). To stabilise penicillins during sample processing, several analytical methods apply derivatisation (benzoic anhydride with 1,2,4-triazole mercury (II) reagent [1214] or formaldehyde with trichloroacetic acid [5]). An additional aim of derivatisation is the introduction of chemical moieties that improve detection by liquid chromatography–ultraviolet or fluorescence [5, 12].

Nowadays, liquid chromatography–tandem mass spectrometry (LC–MS/MS)-based methods for the determination of penicillins are widely available [15]. The reproducibility of quantitative LC–MS/MS methods is often improved when isotopically labelled internal standards can be incorporated. However, the introduction of isotopically labelled analogues of the penicillins is hampered by their complex structure. The only suitable penicillin internal standards described thus far are benzylpenicllin-13C2 [16], benzylpenicillin-d7 [7, 17], benzylpenicillin-d5, ampicillin (AMP)-d6 and nafcillin (NAF)-d6 [18]. Alternatively, several methods describe the use of one of the other penicillins as IS [5, 1923]. The use of other (non-deuterated) penicillins as internal standards is not advisable as these also have veterinary uses and they may exhibit different behaviours during extraction and chromatography as well. So far, no methods have been described that combine derivatisation of penicillins with LC–MS/MS. In this study, the use of piperidine (PIP) to stabilise penicillin residues in matrix by the formation of penicillin–piperidine derivatives is presented. Deuterated piperidine derivatives have been synthesised for all relevant penicillins, enabling the use of isotope dilution for the analysis of penicillins. The development and validation, according to EU guidelines of Commission Decision 2002/657/EC [24], in porcine muscle and kidney tissue as well as in milk and at low carryover levels in feedstuffs is described.

At RIKILT—Institute of Food Safety, feed and porcine tissue samples are screened with an in-house developed microbiological method [25]. Samples suspected to contain ß-lactams are subsequently analysed using the confirmatory and quantitative method described in this paper. The method for animal tissues and milk has been in use for over 5 years and the method for feed for over 3 years.

Experimental

Chemicals and reagents

All chemicals and solvents were at least of analytical reagent grade and were purchased from local commercial suppliers. Acetic acid, ammonium acetate (NH4Ac), phosphoric acid (85%), disodium hydrogen phosphate dibasic salt (Na2HPO4·2H2O) and formic acid (FA, 98–100%) were obtained from Merck (Darmstadt, Germany). Methanol (MeOH) and acetonitrile (MeCN) were supplied by Biosolve (Valkenswaard, the Netherlands), and piperidine was purchased from Fluka (Zwijndrecht, the Netherlands). Piperidine-d11 was obtained from CDN Isotopes (Pointe-Claire, Quebec, Canada). Amoxicillin, ampicillin, penicillin G (PNG), penicillin V (PNV), oxacillin (OXA), cloxacillin (CLO), dicloxacillin (DIC) and nafcillin were supplied by Sigma-Aldrich (Zwijndrecht, the Netherlands). LC quality water (conductivity at least 10 MΩ cm−1) was prepared with a Milli-Q water purification system (Millipore, St. Louis, MO, USA).

Synthesis of penicillin–piperidine derivatives

Derivatised penicillins were prepared by adding a piperidine solution in water (1.00 ml, 200 mM) to the solid penicillin base (0.020 mmol). The mixture was shaken gently until the penicillin was completely dissolved. The reaction was allowed to proceed for 30 min at room temperature, whereupon acetic acid (0.18 ml, 1 M) was added. Finally, NH4Ac buffer (100 mM, pH 6.0) was added to obtain a stock solution with a concentration of 1 mg ml−1 of penicillin base equivalent. The corresponding deuterated analogues were obtained by using a piperidine-d10 solution in water. The derivatised penicillins were stable for 1 year when kept at −20 °C. The purity of the piperidine derivatives was assessed with LC–MS/MS and 1H-nuclear magnetic resonance (NMR) and found to be higher than 90%. No non-derivatised starting material could be detected in the synthesised standards (<1%).

Standard solutions

Stock standard solutions (1 mg ml−1) of penicillin references were prepared in MeOH, taking into account the purity of the reference standard material. The stock solutions were stable for 1 week when stored at 4–8 °C.

From the stocks mixed, penicillin working standard solutions (PWS) were prepared. The concentrations of the individual penicillins were chosen to match the specific MRL in muscle and kidney tissue and in milk. For the analysis in tissue samples, a mixed penicillin working solution (PWS-tissue) was prepared by adding 100 μl of stock solution of AMX, ampicillin (AMP) and PNG; 50 μl of PNV stock; and 600 μl of OXA, CLO, DIC and NAF stock solution to a 100-ml volumetric flask and making up to the mark with water. For milk samples, a mixed working solution (PWS-milk) was prepared by adding 16 μl of AMX, AMP, PNG and PNV stock solution and 120 μl of OXA, CLO, DIC and NAF stock solution to a 100-ml volumetric flask and making up to the mark with water. For feed samples, a mixed working standard (PWS-feed) was prepared by adding 200 μl of all penicillin stock solutions to a 100-ml volumetric flask and making up to the mark with water. PWS were stable for 1 week when stored at 4–8 °C.

Mixed working standard solutions of penicillin-d0-piperidine derivatives (PPWS) and mixed internal standard solutions of the penicillin-d10-piperidine derivatives (PPIS) were prepared from the synthesised standards, matching the specific MRL levels in muscle and kidney tissue and in milk and were mixed in the same proportions as described for the penicillin working solutions. The PPWS and PPIS were stable for 1 month when stored at 4–8 °C.

Working standards for tissue, milk and feed containing mixtures of the penicillin-d0-piperidines and penicillin-d10-piperidines were prepared by adding 250 μl of the specific PPWS and 250 μl of the corresponding PPIS in a 10-ml volumetric flask and making up to the mark with water. These working standards were used to check the performance of the LC–MS/MS system.

Instrumentation

An Agilent 1100 Series high-performance liquid chromatography (HPLC) system (Agilent Technologies, USA) coupled to a Quattro Ultima® tandem mass detector (Waters-Micromass, Manchester, UK), both operating under MassLynx® software, was used for sample analysis. The mass spectrometer was operated in electrospray negative mode, and data acquisition was in multiple reaction monitoring mode (MRM). The precursor/product ions monitored are listed in Table 2. The source settings were as follows: capillary voltage 2.6 kV, source temperature 120 °C, desolvation temperature 300 °C, cone voltage 30 V, cone nitrogen gas flow 180 l h−1, desolvation gas flow 580 l h−1. Argon (2.2 × 10−3 mbar) was used as the collision gas, and the multiplier was operated at 750 V. The collision energy (17–23 eV) was optimised for each analyte and each MRM (Table 2). The HPLC system was equipped with a 3.0 × 150 mm, 5 µm Symmetry C18 column (Waters, Milford, PA, USA) kept at 40 °C. A binary gradient mobile phase was used at a flow rate of 0.4 ml min−1 with solvent A (0.2% FA in water) and solvent B (0.2% FA in MeCN/water 9/1 (v/v)). A splitter was used to reduce the solvent flow to 0.15 ml min−1 entering the mass spectrometer. The gradient started isocratically for 1 min at 15% B, followed by a linear increase to 50% B in 7 min, followed by a linear increase to 70% B in 5 min. After an isocratic hold at 70% B for 1 min, the gradient returned to 15% B in 1 min for equilibration of the column. Total run time was 20 min. Sample injection volume was 50 µl.

Table 2 LC–MS/MS precursor/product ion combinations (quantifier italicised) monitored in ESI negative mode

1H-NMR spectra were recorded on a Bruker Avance 400 MHz spectrometer. Presaturation was performed on the residual HDO resonance. A 90° pulse was used; the total relaxation delay was 3.7 s; the spectral width was 5,000 Hz. The data were acquired in 16 K data points. Before Fourier transformation and phasing, a one-third-shifted sine squared window multiplication was applied and a zero filling to 128 K data points was applied. Calibration of spectra was achieved by setting the HCD2 resonance of deuterated methanol to 3.27 ppm.

Sample preparation: porcine muscle and kidney tissue

Porcine kidney and muscle tissue samples (2 g) were minced and homogenised with phosphate buffer (2 ml, 200 mM, pH = 6.0). Internal standard mixture (PPIS-tissue, 100 µL) was added to all samples. After an incubation time of 30 min at room temperature, phosphate buffer (100 mM, pH = 8.0, 40 ml) was added. The tubes were vortexed for 30 s, followed by addition of piperidine (300 µl). The tubes were vortexed for 30 s, and the pH was checked with a pH stick and if necessary adjusted to 11–12 with piperidine. After agitation for 5 min on a rotary tumbler, phosphoric acid (100 µl) was added and the samples were vortexed; the pH was adjusted to 8.0 ± 0.5 with phosphoric acid if necessary. Next, the samples were heated in a water bath at 85 °C. After 5 min, the samples were quickly cooled to room temperature by using an ice bath. The samples were centrifuged at 2,700×g for 10 min, and the supernatants were filtered through a wad of cotton wool into polypropylene tubes. Cloudy extracts were centrifuged at 15,000×g for 15 min. Oasis® HLB (60-mg, 3-cm3 solid-phase extraction (SPE) cartridges, Waters, Manchester, UK) was conditioned using 2 ml of MeOH, 2 ml of water and 2 ml phosphate buffer (100 mM, pH = 8.0). The cartridges were loaded with extract (10 ml) and subsequently washed with 2 × 2 ml of water. After drying under reduced pressure, cartridges were eluted with 3 ml MeOH/water (80/20, v/v). The eluate was evaporated to dryness at 60 °C under a stream of nitrogen and redissolved in 500 µl 2% MeCN.

Sample preparation: milk

For the analysis of raw bovine milk, the porcine tissue method was used with a few adjustments. To the milk samples (4 g), internal standard mixture (PPIS-milk, 100 µl) was added. Phosphate buffer (100 mM, pH = 8.0, 40 ml) was added, and the tubes were vortexed for 30 s. Piperidine (300 µl) was added, and after vortex agitation for 30 s, the pH was checked with a pH stick and if necessary adjusted to 11–12 with piperidine. After agitation for 5 min on a rotary tumbler, phosphoric acid (100 µl) was added and the samples were vortexed; the pH was adjusted to 8.0 ± 0.5 with phosphoric acid if necessary. The samples were centrifuged at 2,700×g for 10 min, and the supernatants were filtered through a wad of cotton wool into polypropylene tubes. SPE cartridges were loaded with 20 ml of extract, and the final residue was reconstituted in 200 µl of 2% MeCN.

Sample preparation: animal feed

Feeding stuffs were ground to a particle size of 0.5 mm. To the feed samples (2 g), internal standard mixture (PPIS-feed, 200 µl) was added. After an incubation time of 30 min at room temperature, phosphate buffer was added (100 mM, pH = 8.0, 40 ml). The tubes were agitated for 15 min on a rotary tumbler. The samples were centrifuged for 10 min at 2,700×g, and with a pipette 10 ml of the supernatant was transferred to a 14 ml polypropylene tube. Piperidine (300 µl) was added, and the samples were vortexed immediately. After agitation, the pH was checked and if necessary adjusted with piperidine to 11–12. After 5 min, extracts were adjusted to pH 8.0 ± 0.5 with phosphoric acid (100–200 µl). Subsequent purification with Oasis® HLB followed the procedure described for porcine tissue. After evaporation, the residue was reconstituted in 500 µl of water/MeCN/FA (98/2/0.2, v/v/v).

For accurate quantification, feed samples were analysed using a multi-level standard addition approach. The chosen addition levels depend on the concentrations expected in the feed. In case suspect samples (based on the in-house microbiological screening assay) were analysed, the indicative values derived from the screening assay were used to determine standard addition levels. Depending on the expected concentration, six standard addition levels were chosen and spiked accordingly. To samples suspected to contain more than 5 mg kg−1, a larger volume of PPIS mixture (2 ml) was added. Depending on the results from the microbiological screening, an aliquot of the crude extract was taken and diluted with phosphate buffer (100 mM, pH = 8.0). Samples containing less than 1 mg kg−1 were not diluted, whereas samples suspected to contain between 1 and 5 mg kg−1 were diluted five times with phosphate buffer and samples containing between 5 and 50 mg kg−1 were diluted 50 times before derivatisation.

Quality control samples and matrix-matched standard preparation

For tissue and milk samples, calibration was performed using matrix-matched standards (MMS) prepared by adding the appropriate amounts of the penicillin mixed standard solution to blank matrix. For porcine tissue, MMS were prepared corresponding to 0 MRL, 0.25 MRL (25 µl of PWS-tissue), 0.5 MRL (50 µl), one MRL (100 µl), two MRL (200 µl) and five MRL (500 µl). Internal standard mixture (PPIS-tissue, 100 µl) was added to all samples. For milk, MMS were prepared corresponding to 0 MRL, 0.25 MRL (25 µl of PWS-milk), 0.5 MRL (50 µl), one MRL (100 µl), two MRL (200 µl) and five MRL (500 µL). Internal standard mixture (PPIS-milk, 100 µl) was added to all samples. Two blank samples spiked with IS were included in each series for the preparation of matrix-matched recovery standards. Penicillin-d0-piperidine mixed standard solution (PPWS-tissue) was added to the dried extracts to obtain concentrations corresponding to 0.5 MRL (12.5 µL) and two MRL (50 µL) in a total volume of 500 µl 2% MeCN. For milk, PPWS-milk was added to the dried extracts to obtain concentrations corresponding to 0.5 MRL (25 µL) and two MRL (100 µl) in a total volume of 200 µl 2% MeCN.

At the beginning of each series, a working standard containing the penicillin-d0-piperidines and the penicillin-d10-piperidines was injected to check the performance of the instrument.

Validation

The method for porcine muscle tissue was validated in a 5-day experimental setup. On three separate occasions, using three different blank porcine muscle materials, samples were fortified with a mixture of penicillins (PWS-tissue) at three levels (0.5 MRL, one MRL and 1.5 MRL). Six replicates, at each of the three fortification levels, were analysed on each day. Accuracy, intra-day precision (repeatability), inter-day precision (within-lab reproducibility, muscle tissue only), recovery, linearity, CCα and CCß were determined. Additionally (day 4), 20 blank samples were analysed to determine selectivity, specificity and limit of detection (LoD; the concentration at which the quantifier product ion has a signal to noise ratio of 3). The method for muscle tissue was tested (day 5) for ruggedness according to Youden [24]. Variables that were tested included the pH of the phosphate buffer, the piperidine derivatisation time, the heating temperature of the derivatised extract, the drying time of the SPE cartridge before elution of the purified extract and the composition of the elution solvent and the time lag before the dried sample was reconstituted. Porcine kidney tissue was additionally validated using six replicates, at each of the three fortification levels on one occasion.

The validation for milk was limited to seven replicate measurements at MRL on 1 day. Intra-day precision and accuracy at MRL level, linearity, specificity and selectivity were determined.

The validation for feedstuffs was carried out on seven separate occasions using multi-level standard addition. Of each of the 20 blank feed samples (representing feeding stuffs intended for poultry, porcine, bovine, ovine), 11 subsamples were processed. Two samples were not spiked with PWS-feed (blanks), six samples were fortified at one of the validation levels (0.1, 0.2, 0.5 mg kg−1) in duplicate and the remaining three samples were fortified at 1, 2 and 5 mg kg−1, respectively. Each validation level was quantified by using the first three addition levels and applying linear regression of the least squares. The sum of the peak areas of the two product ions was used to calculate the response factor relative to the IS. Accuracy, intra-day precision, inter-day precision, linearity, CCα, CCß, specificity, selectivity and stability of the extracts were determined.

Results and discussion

Synthesis of penicillin–piperidine derivatives

For the antibacterial activity of penicillins, an intact ß-lactam ring is essential. Acid- or base-catalysed hydrolysis of the ß-lactam ring will result in inactive penicilloic acids and degradation products thereof [2628]. Penicillins containing a primary amino group may also isomerise to (inactive) diketopiperazine derivatives [27, 28]. Other nucleophiles, such as alcohols and amines, can also react with the ß-lactam ring, resulting in the formation of penicilloylic esters and (substituted) amides [29, 30]. In case ammonia or simple primary amines are used, the reaction is also called aminolysis. To our knowledge, the reaction of penicillins with secondary amines has not been studied in detail. We found that the reaction of a variety of penicillins with a concentrated solution of piperidine in water proceeded within minutes at room temperature. The synthesis of the derivatised penicillins does not require any special equipment. For the penicillins studied, a single compound was produced in high yield. In all cases, the corresponding piperidine derivative was formed by addition to the ß-lactam carbonyl group, followed by ring opening. In Fig. 2, the 1H-NMR spectrum of amoxicillin-d0-piperidine (AMX-d0-PIP) is shown as an example. In AMX ß-lactam protons, H5 and H6 resonate at 5.40 and 5.58 ppm, while in the ring-opened AMX-d0-PIP they have shifted to 4.85 and 4.92 ppm. The piperidine protons are found at 1.65 ppm (6H) and 2.96 ppm (4H).

Fig. 2
figure 2

1H-NMR of amoxicillin-d0-piperidine, recorded in CD3OD. Stars indicate solvent and residual water resonances

It is possible to isolate the penicillin–piperidine derivatives by crystallisation, but in practise it is more convenient to prepare the standards in situ and store them at −20 °C or lower, without further purification. It was verified by 1H-NMR that during the reaction no or little epimerisation at the 6 position occurs. However, during long-term storage, some epimerisation may occur in solution, even when the standards are kept at −20 °C. After 12 months at −20 °C, a stock solution of AMX-d0-PIP contained 15% of the 6-α epimer. Therefore, it was decided to synthesise new standards every 12 months.

The MS/MS spectra of AMX-d0-PIP and AMX-d10-PIP are presented in Fig. 3. The addition of piperidine is evident from the mass increment of 85 Da in the deprotonated molecular ion (M-H) (AMX: m/z 364). Two major fragments are formed upon collision-induced dissociation: m/z 371 (loss of m/z 78 or H2S + CO2) and m/z 277 (loss of m/z 172, loss of H2S + CO2 + loss of phenol (m/z 94)). Both fragments still contain the piperidine ring. In the MS/MS spectrum of the corresponding d10-analogue, the expected mass shift of 10 Da is seen for the (M-H) and both fragments.

Fig. 3
figure 3

MS/MS fragmentation spectra of amoxicillin-d0-piperidine (top) and amoxicillin-d10-piperidine (below), recorded in electrospray negative mode at a collision energy of 15 eV

Method development

Penicillins do not only react quickly with piperidine but also with other primary (e.g., methylamine, aminoethanol) and secondary amines (diethyl amine, morpholine, piperazide). Piperidine was chosen as the derivatisation reagent of choice because a suitable deuterated analogue is available and the derivatives produced exhibit favourable chromatographic behaviour. Amoxicillin is the most polar penicillin, eluting shortly after the void volume on a C18 HPLC column. AMX-d0-PIP exhibits increased retention, eluting at 4.4 min with the gradient used in this study (Fig. 4). The derivatised penicillins are most conveniently recorded in electrospray negative mode. For all compounds, loss of the m/z 78 fragment is the first or second most prominent fragmentation pathway. Next to the loss of the m/z 78 fragment, for each penicillin derivative, one structure-specific fragment has been included in the MRM method as well (Table 2).

Fig. 4
figure 4

LC–MS/MS chromatograms obtained for a blank porcine muscle sample spiked with a penicillin mix at MRL. For each penicillin–piperidine derivative, the quantifier trace is shown and the chromatogram for the corresponding IS transition. Stars indicate matrix interference

It was verified that the derivatisation is specific for penicillins containing an intact ß-lactam and that no reaction occurs with hydrolysed or degraded penicillins that potentially could interfere with the mass spectrometric analysis. Penicillins were degraded in solution by reaction with 0.1 N sodium hydroxide and with 0.1 N hydrochloric acid, until LC–MS/MS analysis indicated the complete conversion to hydrolysed or degraded products [26, 28]. Piperidine was added to the samples and subsequent LC–MS/MS indicated no change in the composition of the samples, nor any formation of penicillin–piperidine derivatives from the hydrolysed or degraded penicillins. Hence, it can be concluded that the derivatisation is very specific for penicillin residues containing an intact ß-lactam ring and that, in line with EU regulations, only the pharmacologically active residues are determined [4].

The pH of the extract during derivatisation was identified as a critical factor. In Fig. 5, the effect of pH on the derivatisation efficiency is shown. Piperidine was added in three different amounts to a solution of AMX in phosphate buffer (100 mM) with a pH ranging from 3 to 9. By addition of piperidine, which acts as a base, the pH of the buffer solution is raised. Figure 5 shows that derivatisation is catalysed by high pH (11 or higher); at neutral or acidic pH, the reaction proceeds very slowly and by-products (due to hydrolysis) are produced. Therefore, an excess of piperidine is used to insure an alkaline pH. However, prolonged exposure of the extracts to high pH may in turn lead to degradation of the matrix and this may affect sample cleanup. By adding phosphoric acid after 5 min of reaction, degradation of the matrix is minimised.

Fig. 5
figure 5

Derivatisation of amoxicillin (5 μg ml−1) with piperidine as function of the pH of the solution (100 mM phosphate buffer, pH 3 to 9) and the amount of piperidine added (n = 2± SD). The effect of piperidine on the final pH of the solution is indicated

Although the use of deuterated internal standards facilitates the accurate quantification of penicillin residues in animal tissues and milk, the results for animal feed initially were less satisfactory, even with internal standard correction. This may be attributed to the large variability in composition, which can be found in feedstuffs. Raw materials and additives of feeds introduce many interfering components, such as oils, fats, proteins and salts, which can occur at very high levels. These constituents may influence the efficiency of sample preparation by causing differences in derivatisation and extraction efficiency and they may induce ion suppression effects during LC–MS/MS analysis. The internal standards are added to samples as derivatives and therefore only correct for ion suppression and variability in cleanup efficiency after derivatisation, but not for differences in derivatisation efficiency. Multi-level standard addition (MLSA) is therefore required for the analysis of animal feeds. MLSA eliminates differences in derivatisation efficiency, thereby improving accuracy and precision of the determination of penicillins in animal feed [31, 32]. Another possibility to reduce ion suppression effects may be dilution of the extract, as described by de Baere et al. [5], but this method was developed for medicated levels (200–500 mg kg−1). This is not an option when penicillins at a carryover level as low as 0.1 mg kg−1 have to be determined. Feeds with concentrations higher than 5 mg kg−1 need to be diluted to comply with the linear dynamic range of the MS detector, but it was not investigated if MLSA could be skipped for these diluted extracts.

Validation

The validation data for porcine muscle tissue were performed according to CD 2002/657/EC [24], and the results obtained are presented in Table 3. In Fig. 4, LC–MS/MS chromatograms are shown for a processed blank muscle tissue sample spiked with penicillins at MRL. Accuracy in muscle tissue varied from 94% to 113% with an intra-day precision (relative standard deviation (RSD)r, n = 18) between 5.2% and 13.4% and an inter-day precision (RSDRL, n = 18) between 5.5% and 23%. Requirements for accuracy (80–110%), repeatability and within-laboratory reproducibility were met for all compounds at all levels with the exception of the accuracy for CLO (113%, 1.5 MRL). Only a few multi-residue HPLC [13, 14] and LC–MS methods [31] have been reported that include all penicillins described in this paper. These studies reported comparable results. Slightly higher variability was observed for PNV, which at least in part, may be caused by the lower fortification levels, compared to the other penicillins. The CCα and CCß values that were obtained for all penicillins at MRL comply with the EU criteria and were comparable to those reported for bovine muscle tissue, even though only four replicates per level per day (n=12) were used in that study [31]. The average overall recovery determined at 0.5 and two MRL was 70% for AMX and higher than 80% for the other penicillins. No significant differences in recovery were observed between both fortification levels. Linearity of the MMS curve (0.25–5 MRL) was 0.995 or higher on each of the five validation days. Only in two instances (PNV 0.992 and NAF 0.986, both on day 4) was (slightly) lower linearity observed. For each penicillin, the estimated LoD is five times or more below the specific MRL.

Table 3 Validation results obtained for the determination of penicillins in porcine muscle and kidney tissue

The method was tested for ruggedness in tissue according to Youden. From the experiment, it was concluded that samples must be reconstituted immediately after drying. By analysing 20 different blank porcine tissues, the method was proven to be specific and selective. No interfering signals were observed in any of the blanks at the relevant retention times.

Kidney tissue was validated using a single-day experimental setup because of similarity in matrix (porcine tissue) and fortification levels. The validity of this approach was checked by preliminary experiments during method development. These tests indicated that the performance characteristics were likely to be in the same range as the determined parameters for muscle tissue. Accuracy in kidney tissue varied from 83% to 111% with an intra-day precision (RSDr, n = 6) between 4.0% and 17.4% (Table 3). Average recovery in kidney tissue was higher than 80% with the exception of AMX (36%). Lower-than-average recoveries for amoxicillin in kidney tissue were also observed by Becker et al. [31]. They attributed this to loss of retention on SPE by overload of matrix compounds from the extract. Lower recoveries for AMX have also been reported for other matrices such as wastewater and honey [17, 33]. Although recovery for AMX in kidney tissue is significantly lower than in muscle tissue, the accuracies in combination with RSDs obtained indicate that the internal standard corrects adequately for the loss in recovery. No other significant differences were observed between kidney and muscle samples during this study. This is in accordance with the report of Verdon et al. [14], but Becker et al. deemed kidney to be a more difficult matrix [31].

The method for muscle and kidney tissue was adapted for bovine milk. To accommodate for the (approximately tenfold) lower MRLs in milk, the protocol for porcine tissue was adjusted. A larger portion was processed, the final sample extract volume was reduced and the injection volume increased. In this way, the required detection limits could be achieved and confirmation of identity could be established for the penicillins (AMX, AMP, PNV, PNG) with an MRL at 4 µg kg−1. To check the validity of the adjustments of the method made for bovine milk, the method was additionally validated at MRL using a single-day experimental setup. Accuracy in milk varied from 87% to 103% with an intra-day precision (RSDr, n = 7) between 4.9% and 17.5% (Table 4). The accuracies obtained in milk were better than those reported by Kantiani [23]. On the other hand, intra-day precisions were slightly higher than those reported by Sørensen et al. [22] and Becker et al. [31]. Because the method for milk performed well at MRL, no further validation at other concentrations was carried out. LC–MS/MS traces of a compliant and a non-compliant milk sample (containing PNG above the MRL of 4 µg kg−1) from the National Residue Monitoring Program are shown in Fig. 6.

Table 4 Single-day validation results obtained for penicillins in milk (n = 7) at MRL
Fig. 6
figure 6

Selected ion chromatograms for: a blank milk sample spiked with 4 μg kg−1 PNG-d10-PIP IS; b a (non-compliant) milk sample obtained from the National Residue Monitoring Program, containing 5.2 μg l−1 PNG; c blank feed sample spiked with 10 mg kg−1 AMP-d10-PIP IS; d A feed sample containing 45 mg kg−1 AMP (non-compliant). Of the feed extracts, 200 μl was used for further SPE cleanup

As there are no EU criteria established for the validation of veterinary drugs in animal feeds, the method for animal feed was validated according to CD 2002/657/EC for practical purposes. The validation was aimed at low carryover levels (100 to 500 µg kg−1). With the exception of NAF, accuracy in animal feed varied from 88% to 116%, the intra-day precision (RSDr, n = 18) between 11% and 32% and the inter-day precision (RSDRL, n = 18) between 11% and 36% (Table 5). The accuracy for NAF was rather poor, resulting in an overestimation of the concentration (especially at the lowest fortification level) and consequently in a very high CCα and CCß. Inspection of the LC–MS/MS data revealed that in some of the samples a matrix interference was co-eluting with NAF-d0-PIP. By analysing 20 different animal feeds, it was verified that for the other penicillins the method was specific and selective. No interfering signals were observed in any of the blanks at the relevant retention times. The linearity of the MLSA quantification (based on three additional fortification levels) ranged from 0.996 to 1.000. LC–MS/MS traces of a blank animal feed sample and a non-compliant sample containing 45 mg kg−1 AMP are shown in Fig. 6. To our knowledge, a multi-residue method for penicillins at carryover level (0.1–2 mg kg-1) in feed has not been reported before. At medicated levels (200–500 mg kg−1), mostly single-residue methods have been described for AMX (which is the only penicillin registered for use in feed). The RSDr and RSDRL obtained by these methods are invariably lower than the ones reported in this study, but this is to be expected considering the much lower target levels in this study [5, 34, 35]. LoDs are much lower for the method reported in this paper.

Table 5 Validation results obtained for the analysis of penicillins in animal feed

The derivatisation with piperidine was also tested on the other important group of ß-lactam antibiotics, the cephalosporins. Ideally, these compounds together with the penicillins and clavulanate should be incorporated into a general multi-residue ß-lactam method. Derivatisation experiments with clavulanate and several cephalosporins have thus far proved unsuccessful or produced the desired derivatives only with a low yield (results not shown). Apparently, cephalosporins do not react with piperidine in the same straightforward way as the penicillins.

Conclusion

The derivatisation of penicillins with piperidine and the incorporation of the corresponding deuterated penicillin piperidine internal standards not only improved the stability of the penicillins during sample workup but also improved accuracy, repeatability and within-lab reproducibility of the analytical method significantly. The developed multi-residue LC–MS/MS method for the analysis of penicillins in porcine tissue, milk and animal feed has been validated according to Commission Decision 2002/657/EC. Except for the quantitative determination of NAF in animal feed, the obtained validation parameters were all in accordance with the criteria mentioned in the decision. The in-house validation data collected showed that the method is robust and fit for purpose. This was also confirmed by the results of two recent proficiency tests on penicillin residues in milk and in porcine tissue. Excellent results were obtained for penicillin G (Z score −0.26) and cloxacillin (Z score 0.21) in milk [36] and for cloxacillin in muscle tissue (Z score −0.06) [37].

The method has been successfully implemented in the Dutch National Residue Control Program for the analysis of penicillins in kidney, muscle, milk and animal feed samples.